How to maintain and repair ultrasonic cleaning machines

Time:2014-10-14

The cleaning machine consists of an ultrasonic generator and an ultrasonic transducer. The ultrasonic transducer is a sandwich type transducer made of piezoelectric ceramic material, which generates mechanical vibration under the action of an alternating electric field.

 

Common problems with ultrasonic cleaning machine generators include:

⑴ Turn on the power switch of the ultrasonic cleaning machine, but the indicator light does not light up. When repairing this situation, it is necessary to check whether the power switch is intact and whether the leakage switch is closed. If the switch is intact, then check if the fuse is overloaded and blown, which can basically solve the problem.

After turning on the power switch, the indicator light is on, but there is no ultrasonic output. This situation is quite complex. When repairing, first check if the connection plug between the transducer and the ultrasonic power board is loose, and then check if the fuse is blown. If everything is normal, it is possible that there is an internal fault in the ultrasonic power generator. Use a multimeter to open the power cord and check if all the live wires are connected. After troubleshooting the generator, check if the ultrasonic transducer is burnt out and if it needs to be replaced.

The DC fuse of the ultrasonic cleaning machine is blown. It may be due to the burning of the rectifier bridge stack or power tube, or it may be due to aging of the transducer and unstable current, all of which can cause power generator failure. Pay more attention during maintenance.

After turning on the power switch, the ultrasonic cleaning machine produces ultrasonic output, but the cleaning effect is not as ideal as before. This situation is mostly caused by improper use, such as improper cleaning liquid level in the cleaning tank, which is too high or too low; The ultrasonic frequency coordination was not adjusted properly; The liquid temperature in the cleaning tank is too high or too low; Improper selection or insufficient amount of cleaning solution. Of course, it cannot be ruled out that after the transducer is debonded, most of the ultrasonic energy cannot be transmitted to the cleaning solution. In this case, it should be promptly sent back to the manufacturer for repair to avoid burning the oscillator.

Common issues with ultrasonic transducers:

(1) "Ultrasonic transducer is damp". Generally, a megohmmeter is used to check the plug connected to the transducer. By checking the insulation resistance value between the positive and negative poles of the transducer, it can be determined. It is generally required that the insulation resistance be greater than 30 megaohms or more. If the insulation resistance value cannot be reached, it is likely that the transducer is damp. The maintenance method is to place the entire transducer (excluding the spray coated shell) in an oven set at around 100 ℃, dry it for three hours, or use a hair dryer to remove moisture until the resistance value is normal. The transducer oscillator ignites, causing the ceramic material to shatter. During maintenance, a combination of visual inspection and a megohmmeter can be used for inspection. Generally, as an emergency response measure, individual damaged transducers can be disconnected without affecting the normal use of other transducers We know that most manufacturers use adhesive to fix transducers, but ultrasonic cleaning machines may experience debonding due to vibration after long-term use. Some manufacturers also use the method of bonding and tightening with screws, which generally does not cause delamination. Due to the action of the screws, the oscillator will not fall off the vibration surface after delamination. The general judgment method is to gently shake the tail of the oscillator by hand and carefully observe the adhesive situation on the vibration surface to make a judgment.

After the detachment of the general oscillator, the output power of the ultrasonic power supply is normal. However, due to poor connection between the oscillator and the vibration surface, the vibration effect of the vibration surface is not good. After a long time, due to the inability to release energy, it is likely to burn out the oscillator. The detachment of the oscillator is quite troublesome for users to repair, and in general, it can only be sent back to the manufacturer for repair. An effective way to avoid detachment of the oscillator is to avoid hitting the vibrating surface during normal use.

(4) "Vibration surface perforation": Generally, ultrasonic transducers may experience vibration surface perforation after being used at full load for several years. This is due to long-term high-frequency vibration fatigue of the stainless steel plate on the vibration surface. Vibration surface perforation indicates that the service life of the transducer has expired and can only be replaced for maintenance.

   

 

 

 

Disclaimer:The authenticity of this information has not been verified by our website and is for your reference only. The purpose is to convey more information; If there is any error, error, or infringement, please contact us via email( chinavta@szxwds.com )We would greatly appreciate it if you could promptly notify us of the error, error, or infringement, and we will promptly make corrections or deletions.

 

主站蜘蛛池模板: 欧美久久综合九色 | 久久婷婷色一区二区三区 | 91成人在线一区 | 国产欧美日韩aⅴ片免费观看 | 人妻系列专区无码在线 | 深夜福利小视频国产在线观看 | 无遮挡十八禁污 | 欧洲高清视频在线观看亚洲精品 | 97人妻久久精品人人干 | 中日韩视频在线观看 | 刺激性视频黄页网站大全 | 久久人妻无码精 | 国产又黄又大又粗的视频 | 最新免费电影在线免费观看 | 亚洲国产男人的天堂 | 国产成人a视频高清在线观看 | 国产亚洲一区二区手机在线观 | 老司机亚洲精品影院无码 | 国产一区视| 日韩精品成人一区二区在线观看 | 久久精品国产亚洲一区二区三区 | 日韩欧美中文国产在线 | 午夜福利资源大全 | 久久99国产精品成人欧美 | 国产精品免费综合一区视频 | 亚洲va成无码人在线观看天堂 | 精品久久久无 | 人与动欧交视频 | 亚洲av日韩av天堂影片精品 | 国产末成年呦交在线 | 日本免费黄色片 | 国产粉嫩| 久久亚洲精品无码va白人极品 | 高清国产天干天干天干不 | 97国产精品人人爽人人做 | 亚洲综合中文字幕第一页 | 92国产精品午夜福利免费 | 在线观看av永久无码精品成人 | 亚洲精品国品乱码久久久久 | 国产更新国产最新国产主播 | 天天干天天射天天操 |